| Model Number |
M00509011 |
| Medical Device Problem Codes |
Break (1069); Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6 (device codes): imdrf device code a0401 captures the reportable event of loop wire on end of peg detached.Block h10: the returned endovive safety peg kit push method was analyzed.During visual inspection, it was observed that the loop wire section at the distal end of the feeding tube was detached.Therefore, the reported complaint is confirmed.Based on the condition of the returned device, engineers determined that the problem observed most likely occurred due to some technique used during procedure.An excess of force applied could have contributed to the detachment of the distal loop.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on event which led to the reported event.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an endovive safety peg kit push method was used during a percutaneous endoscopic gastrostomy placement procedure performed on (b)(6), 2023.No further information has been obtained despite good faith efforts.This event has been deemed a reportable event based on the investigation finding of a loop wire distal end section of the feeding tube was detached.Please see block h10 for full investigation details.
|
| |
|