| Catalog Number |
02P36-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A literature article by talia wall, ¿a case of a false positive hiv antigen/antibody screen in a pregnant woman at delivery and the clinical importance of reviewing signal-to-cutoff ratio values¿, idcases 33:e01819 (june 15, 2023), presented a case of false repeat reactive architect hiv ag/ab combo results on a 25-year-old woman who came to the labor and delivery unit for a scheduled induction of labor due to large fetal size at a gestational age of 40 weeks 0 day.The patient did not have documented third trimester result for hiv and was therefore screened upon presentation for labor and delivery.The patient was considered low risk of contracting hiv since there were three documented nonreactive hiv screening tests over the course of her previous pregnancies.Pitocin was initiated and a few hours later the architect hiv ag/ab screening test revealed a repeatedly reactive result.Since the likelihood of previous hiv exposure in this patient was low, the medical team assumed a possible laboratory error and continued the augmentation of labor with pitocin.Per the hospital protocol, a fresh blood sample was collected from the patient and a second architect hiv ag/ab combo screening test was performed and the result was repeat reactive.Both samples were then screened with a confirmatory hiv1/hiv2 antibody differentiation immunoassay.The samples were negative for hiv antibodies.The samples were then sent out for confirmatory viral load pcr testing.The patient was presumed to be hiv positive, and pitocin was held with the hopes of slowing down the labor.Zidovudine infusion was initiated, at a dose of 2 mg/kg for the first hour, and 1 mg/kg/h thereafter.The one-hour intravenous loading dose began at 6:30 a.M.(the patient subsequently was given one dose of raltegravir 400 mg po postdelivery at 10 a.M.).The infant was delivered safely via vaginal delivery.After completion of the initial zidovudine infusion, art was held postdelivery for the mother pending viral load results.The mother was also given a dose of prophylactic cefazolin every 8 h post-delivery to address her possible immunocompromised state.The results for both samples sent for confirmatory testing were available at 2 pm on the day following delivery.It was confirmed that no viral rna was detected by the hiv rna pcr test in either sample.The patient was discharged home with the infant.The patient returned for her postpartum visit four weeks later, the screening test, antibody differentiation immunoassay, and viral load were all redrawn with the same results, a reactive screening test, negative antibody immunoassay, and no viral rna detected upon confirmatory testing.Four months after delivery, the patient was finally nonreactive with the architect hiv ag/ab combo test.The article noted that aside from multiple pregnancies, the patient had no other significant medical history.However, the patient she did receive a covid-19 booster vaccine just weeks prior to delivery.No further impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
A literature article by talia wall, ¿a case of a false positive hiv antigen/antibody screen in a pregnant woman at delivery and the clinical importance of reviewing signal-to-cutoff ratio values¿, idcases 33:e01819 (june 15, 2023), presented a case of false repeat reactive architect hiv ag/ab combo results on a 25-year-old woman who came to the labor and delivery unit for a scheduled induction of labor due to large fetal size at a gestational age of 40 weeks 0 day.The patient did not have documented third trimester result for hiv and was therefore screened upon presentation for labor and delivery.The patient was considered low risk of contracting hiv since there were three documented nonreactive hiv screening tests over the course of her previous pregnancies.Pitocin was initiated and a few hours later the architect hiv ag/ab screening test revealed a repeatedly reactive result.Since the likelihood of previous hiv exposure in this patient was low, the medical team assumed a possible laboratory error and continued the augmentation of labor with pitocin.Per the hospital protocol, a fresh blood sample was collected from the patient and a second architect hiv ag/ab combo screening test was performed and the result was repeat reactive.Both samples were then screened with a confirmatory hiv1/hiv2 antibody differentiation immunoassay.The samples were negative for hiv antibodies.The samples were then sent out for confirmatory viral load pcr testing.The patient was presumed to be hiv positive, and pitocin was held with the hopes of slowing down the labor.Zidovudine infusion was initiated, at a dose of 2 mg/kg for the first hour, and 1 mg/kg/h thereafter.The one-hour intravenous loading dose began at 6:30 a.M.(the patient subsequently was given one dose of raltegravir 400 mg po postdelivery at 10 a.M.).The infant was delivered safely via vaginal delivery.After completion of the initial zidovudine infusion, art was held postdelivery for the mother pending viral load results.The mother was also given a dose of prophylactic cefazolin every 8 h post-delivery to address her possible immunocompromised state.The results for both samples sent for confirmatory testing were available at 2 pm on the day following delivery.It was confirmed that no viral rna was detected by the hiv rna pcr test in either sample.The patient was discharged home with the infant.The patient returned for her postpartum visit four weeks later, the screening test, antibody differentiation immunoassay, and viral load were all redrawn with the same results, a reactive screening test, negative antibody immunoassay, and no viral rna detected upon confirmatory testing.Four months after delivery, the patient was finally nonreactive with the architect hiv ag/ab combo test.The article noted that aside from multiple pregnancies, the patient had no other significant medical history.However, the patient she did receive a covid-19 booster vaccine just weeks prior to delivery.No further impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint investigation false reactive architect hiv ag/ab combo results included a review of data and information from the article ¿a case of a false positive hiv antigen/antibody screen in a pregnant woman at delivery and the clinical importance of reviewing signal-to-cutoff ratio values¿, ticket trending review, device history record review, field data review, and labeling review.Return testing was not completed as returns were not available.Data and information provided in the article were reviewed and support the complaint issue.Trending review did not identify any trends for the issue for the product.Device history record review did not identify any non-conformances or deviations with the product list number and complaint issue.The overall performance of the architect hiv ag/ab combo reagent was reviewed using data gathered from customers worldwide.Standard deviation to cutoff and median values of the negative population generated with the product are within the +/-3 sd control limits showing that performance is as expected.Labeling was reviewed and sufficiently addresses the customer's issue.Based on the review of this issue, this case represents a use error as the patient was assumed to be hiv positive and the decision to give hiv treatment to the mother was based on reactive architect hiv ag/ab combo results that were not confirmed accordingly as per cdc recommendations as well as per the architect hiv ag/ab assay package insert/instructions for use.Based on this investigation, no systemic issue or deficiency was identified with the architect hiv ag/ab combo reagent, lot number unknown.
|
| |
|