Two opened knives, in a bp tray were received for the report of knife was not sharp enough during surgery.Samples were visually inspected and found to be nonconforming, sample # 1 - blade tip was observed to be broken off and sample # 2 - blade was observed to have a damaged cutting edge and bent tip.Penetration testing could not be performed due to the damage of the samples.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.The exact root cause could not be determined from the investigation performed.The damage to the returned samples is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when product is improperly removed or inserted after use, or improper handling, or contact with another instrument during surgery or set-up.The damage seen on the returned complaint sample could have contributed to the customers reported issue of dull blade.The exact root cause for the damaged knife samples is unknown, therefore specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Any nonconformance, such as the damaged cutting edge and bent tip exhibited on the returned opened samples, is removed from the lot and scrapped.Functional penetration testing is performed and monitored during the finishing process to ensure the sharpness of the product.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|