Investigation: visual inspection: during the investigation, deposits on the device was determined.Permeability test: a permeability test has indicated that the m.Blue has a blockage.Computer controlled test: due to the blockage, no measurement could be determined.Adjustment test: the m.Blue valve was tested and is not adjustable throughout the normal range.Braking force and brake function test: the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection : after dismantling of the valve, deposits were found in m.Blue.Results: based on our investigation results, we can determine a blockage and adjustment difficulties in the valve.The determined deposits can be named as the cause for the blockage and adjustment difficulties.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|