| Catalog Number |
EIC4845-01 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during ent surgery, the electrode of the reflex ultra 45 coblator was broken in the patient's tissue.Afterwards, local anesthesia was performed for surgery, the tip was removed, and the patient was cured and discharged.The procedure was completed without surgical delay using a smith and nephew back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.Visual inspection observed the returned instrument shows no manufacturing abnormalities.The electrode has been used.Bio debris is present.The electrode is not visible but the tip is packed with bio debris to the edge of the tube.The device was returned in original packaging with the batch number of the complaint on the label.A functional evaluation showed the opened device was tested and plugged into the controller and registered settings (4,2).The wand produced plasma that could be seen at the tip.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include: (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|