|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 1 year since the subject device was manufactured.Based on the results of the investigation, force was applied to the distal end of the h electrode during cleaning of the tissue pad using gauze resulting in damage.The event can be prevented by following the instructions for use which state: ¿ in the unlikely event that there are cracks, hangnails, protrusions, scratches, cracks, or deformations on the probe tip, gripping section, tissue pad, insertion section, thunderbeat transducer exterior, transducer cord, transducer plug, etc., never use the device and replace it with a new spare device, as this may lead to output abnormalities, burns due to high-frequency leakage current, or the tip of the probe falling off.¿ should any crack, scratch, deformation, split, protrusion, or partial separating be observed on the probe tip, grasping section, tissue pad, insertion section, the surface of the transducer, transducer cord, or transducer plug, do not use them and replace the damaged instrument or the transducer with a spare.Using a damaged device may cause burns due to abnormal output or high-frequency (rf bipolar) current leakage or breakage of the probe tip, the tissue pad, and the grasping section.¿ during surgery, remove any dirt such as exposed metal parts on and around the gripping part, or burnt spots on the tip of the probe with a soft object such as sterilized gauze or a soft brush.If you try to scrape it off with a hard object such as a scalpel blade or tweezers, the grip and tip of the probe may be scratched, which may break during the procedure and fall into the body cavity.Additionally, the insulation structure may be damaged and high-frequency leakage currents may cause tissue burns.¿ when cleaning the grasping section or the probe tip during treatment, wipe it with a soft object such as a piece of gauze or a brush if a body fluid or tissue gets burnt onto it.Do not attempt to scrape it with a hard object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, the metal-exposed area around it, the tissue pad, a coated surface, or the probe tip may be scratched and damaged, which may lead to the damaged part falling off inside the body cavity.Olympus will continue to monitor field performance for this device.
|