| Model Number |
VTICMO12.1 |
| Medical Device Problem Codes |
Off-Label Use (1494); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - device code: 1494 - this lens model is contraindicated for patients with age less than that of 21 years.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 12.1mm vticmo12.1 implantable collamer lens, -15.50/+2.5/092 (sphere/cylinder/axis), in the patients left eye (os), on (b)(6) 2023.The lens was explanted on (b)(6) 2023 due to low vaulting and the problem was resolved.The cause of the event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h3 - type of investigation code: 10- device evaluation: the lens was returned in a microcentrifuge vial with liquid.Visual inspection found no visible damage.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|