| Model Number |
2760 |
| Medical Device Problem Code |
Failure to Zero (1683)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device needle turned fully when deflated.An encore 26 advantage kit was selected for use.During deflation, it was noted that the needle meter turns fully when deflated and would not return back to zero.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.Visual inspection was performed.The gauge needle could not be returned to zero, confirming the reported event.No further visual issues were noted with the device.Further functional testing could not be performed because the gauge needle could not be returned to zero.
|
| |
|
Event or Problem Description
|
|
It was reported that the device needle turned fully when deflated.An encore 26 advantage kit was selected for use.During deflation, it was noted that the needle meter turns fully when deflated and would not return back to zero.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|