It was reported that a patient underwent a total hip arthroplasty procedure on (b)(6 2023 and barbed suture was used.The fixation tab was broken during use on joint capsule.The dislodged fixation tab remained intramuscularly, but since it was an absorbable material, the surgeon decided to leave it in place, opened another product, and completed the surgery.There were no adverse consequences to the patient.No further information was provided.
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002 ¿ device not returned.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.
|
|
Product complaint # (b)(4).Date sent to the fda: 10/23/2023.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that it received one needle-suture outside its packaging.During the visual assessment of the sample, it was noted that the swage and attachment area were as expected.Marks on the body needle that appears to be by a surgical instrument could be observed.The suture was examined, body fluids and damage at some barbs were noted and both sides of the fixation tabs were found to be broken.It should be noted that a 100% inspection is performed via the automotive vision system to ensure the tab is present and complete during manufacturing.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one packet was received for evaluation.Upon initial inspection, of the sample, no external damages were observed on the packet.In order to evaluate the conditions of the returned sample, the sample was opened, and the needle-suture combination was placed correctly.The suture was dispensed without problems and the fixation feature was present.In addition, the strand was examined along the strand and no defects, or damage on the barbs could be observed.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Product complaint # (b)(4).Date sent to the fda: 10/23/2023.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|