Model Number 24672
Medical Device Problem Codes Break (1069); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 - initial reporter phone: (b)(6).
 
Event or Problem Description
It was reported that a balloon rupture and catheter shaft break occurred.A 7.0 x 120, 75cm mustang balloon catheter was advanced for dilation.After an unknown number of inflations, the balloon burst.The device was removed from the patient, and upon retrieval, the catheter shaft fractured distal to the balloon.The detached portion was removed by removing the entire guidewire system.The procedure was completed with another of the same device.There were no reported patient complications.
 
Additional Manufacturer Narrative
E1 - initial reporter phone: (b)(6).Corrected field: b5 - describe event or problem: additional event details were provided on (b)(6) 2023 that were not included in the initial report submission on (b)(6) 2023.
 
Event or Problem Description
It was reported that a balloon rupture and catheter shaft break occurred.A 7.0 x 120, 75cm mustang balloon catheter was advanced for dilation.After an unknown number of inflations, the balloon burst.The device was removed from the patient, and upon retrieval, the catheter shaft fractured distal to the balloon.The detached portion was removed by removing the entire guidewire system.The procedure was completed with another of the same device.There were no reported patient complications.It was further reported that prior to insertion of the mustang balloon, the lesion was pre-dilated using an unknown 4-5mm balloon.The mustang balloon was inflated to approximately 10-14atm prior to bursting.Additionally, approximately 35cm of the catheter shaft fractured within the patient.
 
Event or Problem Description
It was reported that a balloon rupture and catheter shaft break occurred.A 7.0 x 120, 75cm mustang balloon catheter was advanced for dilation.After an unknown number of inflations, the balloon burst.The device was removed from the patient, and upon retrieval, the catheter shaft fractured distal to the balloon.The detached portion was removed by removing the entire guidewire system.The procedure was completed with another of the same device.There were no reported patient complications.It was further reported that prior to insertion of the mustang balloon, the lesion was pre-dilated using an unknown 4-5mm balloon.The mustang balloon was inflated to approximately 10-14atm prior to bursting.Additionally, approximately 35cm of the catheter shaft fractured within the patient.
 
Additional Manufacturer Narrative
Device analysis: the device was received was received in two sections due to a complete separation of the shaft.A visual examination identified that the balloon was not folded which indicates that it was subjected to positive pressure.A complete balloon circumferential tear was identified located approximately 2mm distal of the proximal balloon bond.From the complete circumferential tear, the distal section of balloon material was also torn longitudinally along its entire length.A visual and tactile examination of the shaft found a complete separation of the shaft located approximately 2mm distal of the proximal balloon bond.Severe stretching damage and kinking was also noted to the distal section of shaft.The distal marker band was displaced due to the shaft damage and the proximal marker band was not returned for analysis.No issues were noted with the tip of the device.No other issues were identified during the product analysis.Images were provided by the customer depicting the damaged device.E1 - initial reporter phone: (b)(6).Corrected field: b5 - describe event or problem: additional event details were provided on 30aug2023 that were not included in the initial report submission on 21sep2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17793595
Report Number2124215-2023-47493
Device Sequence Number15314278
Product Code FGE
UDI-Device Identifier08714729794424
UDI-Public08714729794424
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/20/2026
Device Model Number24672
Device Catalogue Number24672
Device Lot Number0031090136
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/29/2023
Supplement Date Received by Manufacturer08/30/2023
11/28/2023
Initial Report FDA Received Date09/21/2023
Supplement Report FDA Received Date09/29/2023
12/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1