AOMORI OLYMPUS CO., LTD. SINGLE USE SPHINCTEROTOME V (DISTAL WIREGUIDED); SINGLE USE PRELOADED SPHINCTEROTOME Back to Search Results
Model Number KD-VC411Q-0320
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was able to be confirmed.During the device evaluation it was observed that the wire coating was torn, and the broken part was burnt and melted.When the outer diameter of the knife wire was measured, no abnormalities were found.Also, no problems with the length of the knife wire or other defects were found.This report is also being submitted to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of knife wire breakage could not be determined.Considerations ¿ based on the result of confirmation of the device, and the result of past similar complaint investigation, a likely mechanism causing the cutting wire breakage might be the following: 1.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.2.Under circumstances an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.A likely factor causing tears of the coated portion of the cutting wire might be the following.However, the exact cause could not be determined: 1.It is possible that the slider was slightly pushed causing the cutting wire to deflect.The coated portion of the cutting wire has possibly been rubbed due to the effect of contacting a metal area of the endoscope.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer reported to olympus the single use sphincterotome v (distal wireguided) the knife wire was cut at the timing of energization.The reported issue occurred during a therapeutic biliary endoscopic sphincterotomy (est) procedure.The procedure was subsequently completed with a similar device.There was no patient/user harm or injury reported due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE SPHINCTEROTOME V (DISTAL WIREGUIDED)
Common Device Name
SINGLE USE PRELOADED SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17793934
Report Number9614641-2023-01369
Device Sequence Number6252751
Product Code KNS
UDI-Device Identifier04953170371073
UDI-Public04953170371073
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141991
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-VC411Q-0320
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2023
Initial Report FDA Received Date09/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-