• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RADIUS-7 WI-FI; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RADIUS-7 WI-FI; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 26971
Medical Device Problem Code Sensing Intermittently (1558)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported the radius-7 module "intermittently cuts out." there were no patient impact or consequences reported.
 
Event or Problem Description
The customer reported the radius-7 module "intermittently cuts out."  there were no patient impact or consequences reported.
 
Additional Manufacturer Narrative
Additional manufacuring narrative: other, other text: the returned radius-7 was evaluated.The instrument module lost connection with the battery module with minimal physical manipulation.The top cover of the instrument module was lifted and caused the disruption in the connection of the radius-7 module halves.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIUS-7 WI-FI
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
tahereh sedighi
15750 alton pkwy
irvine, CA 92618
9492977862
MDR Report Key17795170
Report Number3019388613-2023-00233
Device Sequence Number2149349
Product Code MWI
UDI-Device Identifier00843997014106
UDI-Public00843997014106
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K153225
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative,Distributor
Initial Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26971
Device Catalogue Number9903
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2023
Supplement Date Received by Manufacturer10/12/2023
Initial Report FDA Received Date09/21/2023
Supplement Report FDA Received Date11/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-