• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXGUARD EXTENSION SET WITH NEEDLELESS Y-SITE(S), STOPCOCK(S) AND MANIFOLD; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MAXGUARD EXTENSION SET WITH NEEDLELESS Y-SITE(S), STOPCOCK(S) AND MANIFOLD; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MX4436
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.The date received by manufacturer has been used for this field.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd maxguard extension set with needleless y-site(s), stopcock(s) and manifold tubing clamp was missing.The following information was received by the initial reporter with the verbatim: reported issue: the roller clamp on the tubing was not present.
 
Event or Problem Description
It was reported that bd maxguard extension set with needleless y-site(s), stopcock(s) and manifold tubing clamp was missing.The following information was received by the initial reporter with the verbatim: reported issue: the roller clamp on the tubing was not present.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 14-sep-2023.H6: investigation summary a complaint of a set missing a roller clamp was received from the customer.A sample was returned for investigation.Through visual inspection, the customer complaint was confirmed.The set was assembled without a roller clamp.A device history record review for model mx4436 lot number 22119025 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The manufacturing plant was notified of this defect and a review of the manufacturing process was completed.After investigation, the potential root cause of missing roller clamp could be related to the assembler not using the fixture due to it being damaged/inverted.A work order ((b)(4)) was released to carry out an inspection of fixtures to identify the damages and fix them.A quality alert was also created and communicated to the production personnel to reinforce the escalation process and notify when a fixture is damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXGUARD EXTENSION SET WITH NEEDLELESS Y-SITE(S), STOPCOCK(S) AND MANIFOLD
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17795499
Report Number9616066-2023-01927
Device Sequence Number6716793
Product Code FPA
UDI-Device Identifier50885403237476
UDI-Public(01)50885403237476
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMX4436
Device Lot Number22119025
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2023
Supplement Date Received by Manufacturer10/18/2023
Initial Report FDA Received Date09/21/2023
Supplement Report FDA Received Date10/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-