| Catalog Number |
324909 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd veo¿ insulin syringes with bd ultra-fine¿ 6mm x 31g needle leakage occurred.There was no report of patient impact.The following information was provided by the initial reporter: consumer reported found 1 syringe with a liquid coming out the needle before use.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd veo¿ insulin syringes with bd ultra-fine¿ 6mm x 31g needle leakage occurred.There was no report of patient impact.The following information was provided by the initial reporter: consumer reported found 1 syringe with a liquid coming out the needle before use.
|
| |
|