| Catalog Number |
72201492 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a meniscus repair surgery, the t1 of the fast fix was placed in the meniscus tear, but t2 could not be deployed.T1 & t2 were removed from the patient using a grasper.The procedure was successfully completed with non-significant surgical delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was received for evaluation.A visual inspection revealed the device was returned in original packaging with the batch number of the complaint on the label.The t1, t2, and suture were returned off the needle.The knot has been tightened down.The variable depth straw was not returned.Bio debris is present.A functional evaluation could not be performed since the implants are no longer on the device.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include failure to fully actuate the device during implantation.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|