MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE¿ IMPLANT 3.75 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE¿ IMPLANT 3.75 X 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number OSS310
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The doctor reports that the dental implant located in position number #46 failed because it fractured at the head.The doctor reported that the implant broke 4mm below the bone crest.The doctor reports that the procedure was concluded by placing another implant.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number: (b)(4).D4: additional device information unknown / not provided.H4: device manufacturer date unknown / not provided.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Zimvie received one (1) oss310, (osseotite implant 3.75 x 10mm) for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, apparent bone / tissue attached to external threads.The implant's collar was fractured.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.This complaint refers to the specific device being investigated for this complaint record.Device history record (dhr) review was completed for the subject lot number 2018010207.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2018010207 for similar events and no other complaint was identified.Review completed utilizing keywords: fracture implant.Based on the investigation and risk management file review, the most likely root cause determined from the investigation is patient factors.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE¿ IMPLANT 3.75 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key17796744
Report Number0001038806-2023-01826
Device Sequence Number20470798
Product Code DZE
UDI-Device Identifier00844868013105
UDI-Public(01)00844868013105(17)230118(10)2018010207
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K063286
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/07/2023
Device Catalogue NumberOSS310
Device Lot Number2018010207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/22/2023
Supplement Report FDA Received Date03/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age49 YR
Patient SexFemale
Patient Weight75 KG
Patient EthnicityNon Hispanic
-
-