This report is being submitted to report additional information.Zimvie received one implant for evaluation.Visual evaluation was performed, there was signs of use with bone debris on the external threads.No malfunctions detected that could lead to reported event.Device was measured with a caliper and matched print specifications.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Passed sterilization.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.A definitive root cause could not be determined.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|