Model Number 701218150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
The needle support broke, causing damage to the patient and dissatisfaction for the physician.
 
Additional Manufacturer Narrative
No sample was available for evaluation.However, images were provided that showed the devices in pieces and the complaint was confirmed.Argon medical devices has received other complaints that the supercore semi-automatic biopsy instrument is coming apart during or prior to use.Argon has conducted an internal investigation and tracked the affected parts to a narrow time frame resulting from a specific manufacturing event.The plastic housing and plunger can be separated more easily than normal for the lots manufactured during this time frame.Capa c-2021-061 was filed.Capa 2021-061 has been opened to document our investigation into the cause of this problem and the corrective action that are being taken to ensure that this never happens again.To ensure continued customer satisfaction, argon medical devices has decided to issue a voluntary recall of the affected lots because of the high rate of reports of unintentional disassembly of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERCORE BIOPSY INSTRUMENT WITH CO-AXIAL INTRODUCER NEEDLE
Common Device Name
SUPERCORE
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key17798309
Report Number0001417485-2023-00002
Device Sequence Number12523074
Product Code KNW
UDI-Device Identifier00886333005987
UDI-Public00886333005987
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K974814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number701218150
Device Lot Number11406221
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2023
Initial Report FDA Received Date09/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1625425-09/30/2021-001-R
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-