• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. SIMPLYGO; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. SIMPLYGO; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1068987
Medical Device Problem Codes Failure to Power Up (1476); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
A simplygo portable oxygen concentrator was returned to the manufacturer for service and an allegation of thermal pca/no power condition.There was allegation of the device burned and melted at the pb board.No void or gaps were found in the enclosure of the device.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, the allegation of a thermal pca/no power condition was confirmed.Pca component q61 (battery charging circuit) has shorted to the ground and vented from an over current.A visual inspection of the device found missing enclosure fasteners, evidence of water ingress, dirt/dust/biological dander throughout the unit/cooling path, physical damage to the main pca (components bent 90 degrees to designed orientation), and thermal creep isolated to component and lands only.The exterior battery failed due to cell imbalance and the battery charging circuit fusible link opened.The investigation additionally found there was lack of maintenance, evidence of a dropped condition, liquid ingress, and tampering.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17798547
Report Number2518422-2023-24091
Device Sequence Number20274097
Product Code CAW
UDI-Device Identifier00606959005723
UDI-Public00606959005723
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1068987
Device Catalogue Number1068987
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2023
Initial Report FDA Received Date09/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-