| Catalog Number |
209999 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(b)(6) : mps reported difficulty entering into and staying inside haptic boundaries + arm drifting without trigger pull.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding arm haptic issue involving a mako robotic arm was reported.The event was not confirmed.Method & results: -product evaluation and results: the field service engineer reported: problem reproduced: no.Troubleshooting notes: none.Work performed: thorough investigation was done and no issues were found with the system.A full pm was completed to ensure system reliability.Robot is ready for clinical use.Work order disposition: system investigation completed successfully as per service manual.All system checks and tests passed, system is ready for use.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other complaints with similar event(s) for the lot referenced.Conclusions: the alleged failure mode was not confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|