| Model Number |
83786 |
| Medical Device Problem Codes |
Premature Activation (1484); Difficult to Remove (1528); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.The main coil returned inside in a non-bsci device.Visual and microscope inspection revealed that the main coil was stretched, detached and bent at the coil arm section.No more damages can be observed on the device.The functional test could not be performed due only the main coil was returned for the analysis.Dimensional inspection in the main coil revealed that the zap tip od (outer diameter) and primary coil od were within specification.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 04sep2023.It was reported that the coil could not be advance halfway.The target lesion was located in the abdominal cavity.A 12mm x 20cm interlock-35 was selected for use.The device was continuously flush before introduction of the coil.During the procedure, it was noted that the coil could not be advance and withdrawn halfway even after several attempts.The whole catheter and the coil were pulled out together from the patient's body.It was then found that the coil had been detached inside the catheter.The procedure was completed with another of same device.No complications were reported, and patient was stable post procedure.However, device analysis revealed that the main coil detached at the coil arm section.
|
| |
|