• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA SURGICAL OPTICS, INC. 251; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA SURGICAL OPTICS, INC. 251; INTRAOCULAR LENS Back to Search Results
Model Number 251 (+ 21.50 D)
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/05/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This initial emdr is being submitted to fda for an event that occurred outside of the usa.Haptic damage is indicated as a potential malfunction related to the iol, as covered under the warnings section of the product's instructions for use (ifu).251: pkg-18-177 01 en2.Ms2.150151250251.20180401_150,151,250,251.Regarding section h6 - manufacturer's codes for: type of investigation, findings, and conclusion are pending device return and completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
 
Event or Problem Description
Event occurred in new zealand, damaged haptic after implantation.Imdrf code: a0414, material split cut or torn.Product replaced with another lens immediately during surgery.Iol wasexplanted on (b)(6) 2023.
 
Event or Problem Description
Event occurred in new zealand.Damaged haptic after implantation.Imdrf code: a0414, material split cut or torn.Product replaced with another lens immediately during surgery.Iol was explanted on (b)(6) 2023.
 
Additional Manufacturer Narrative
This follow-up report #1 emdr is being submitted to the fda for a reportable event that occurred outside the usa.The report includes additional information not available/included in the initial report.Additional information: g6 - type of report - noted as follow-up #1.H2 - type of follow-up - noted for additional information.H6 - added codes for manufacturer's investigation: type; findings; and conclusion.The product was not returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.No abnormalities were found in the production and inspection records of the product.(serial no.: (b)(6) ; model: 25100b).From our investigation, we couldn't confirm the reported event.No root cause was determined based on the findings of the investigation result.However, based on the available information and our investigation, we believe this event was not caused by our product quality.Capa-23-0017 has been initiated for damaged haptics complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
251
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
HOYA SURGICAL OPTICS, INC.
110 progress
suite 175
irvine CA 92618
Manufacturer (Section G)
HOYA MEDICAL SINGAPORE PTE LTD
455a jalan ahmad ibrahim
singapore, singapore 63993 9
SN   639939
Manufacturer Contact
goutham pendyala
110 progress
suite 175
irvine, CA 92618
9093896317
MDR Report Key17802315
Report Number3006723646-2023-00432
Device Sequence Number12501093
Product Code HQL
UDI-Device Identifier04547480450322
UDI-Public04547480450322
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
P080004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model Number251 (+ 21.50 D)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2023
Supplement Date Received by Manufacturer09/12/2023
Initial Report FDA Received Date09/22/2023
Supplement Report FDA Received Date02/01/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-