| Model Number |
251 (+ 21.50 D) |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This initial emdr is being submitted to fda for an event that occurred outside of the usa.Haptic damage is indicated as a potential malfunction related to the iol, as covered under the warnings section of the product's instructions for use (ifu).251: pkg-18-177 01 en2.Ms2.150151250251.20180401_150,151,250,251.Regarding section h6 - manufacturer's codes for: type of investigation, findings, and conclusion are pending device return and completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
|
| |
|
Event or Problem Description
|
|
Event occurred in new zealand, damaged haptic after implantation.Imdrf code: a0414, material split cut or torn.Product replaced with another lens immediately during surgery.Iol wasexplanted on (b)(6) 2023.
|
| |
|
Event or Problem Description
|
|
Event occurred in new zealand.Damaged haptic after implantation.Imdrf code: a0414, material split cut or torn.Product replaced with another lens immediately during surgery.Iol was explanted on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report #1 emdr is being submitted to the fda for a reportable event that occurred outside the usa.The report includes additional information not available/included in the initial report.Additional information: g6 - type of report - noted as follow-up #1.H2 - type of follow-up - noted for additional information.H6 - added codes for manufacturer's investigation: type; findings; and conclusion.The product was not returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.No abnormalities were found in the production and inspection records of the product.(serial no.: (b)(6) ; model: 25100b).From our investigation, we couldn't confirm the reported event.No root cause was determined based on the findings of the investigation result.However, based on the available information and our investigation, we believe this event was not caused by our product quality.Capa-23-0017 has been initiated for damaged haptics complaints.
|
| |
|