• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", BIPOLAR, 5 X 330 MM, GRASPING FORCEPS, FENESTRATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", BIPOLAR, 5 X 330 MM, GRASPING FORCEPS, FENESTRATED Back to Search Results
Model Number WA64120C
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the insulation at the distal end of his olympus jaws grasping forceps was found torn on several units.There was no patient harm reported as a result of these events.This report is 7 of 8.Please see reports with the following patient identifiers for related events: (b)(6).
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.If additional information becomes available prior to the conclusion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the customer's complaint was confirmed.Device evaluation found the insulation body between the jaws and yoke is worn due to use.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event is likely due to either a quality issue or the reprocessing method and/or the service life.The root cause was unable to be determined.The event can be detected by following the instructions for use (ifu) sections which states: before use, the product must be checked for the following (see also ifu, chapter 4.2.2 ¿specific inspection¿): any type of deformation, dents, cracks, deep scratches, defects in and at the insulation, dirt, corrosion, and labelling visibility.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWS "HIQ+", BIPOLAR, 5 X 330 MM, GRASPING FORCEPS, FENESTRATED
Common Device Name
FORCEPS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17803324
Report Number9610773-2023-02674
Device Sequence Number12435383
Product Code GEI
UDI-Device Identifier04042761062411
UDI-Public04042761062411
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K984417
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA64120C
Device Lot Number23206-0020
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2023
Supplement Date Received by Manufacturer01/30/2024
Initial Report FDA Received Date09/22/2023
Supplement Report FDA Received Date02/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-