This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the customer's complaint was confirmed.Device evaluation found the insulation body between the jaws and yoke is worn due to use.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event is likely due to either a quality issue or the reprocessing method and/or the service life.The root cause was unable to be determined.The event can be detected by following the instructions for use (ifu) sections which states: before use, the product must be checked for the following (see also ifu, chapter 4.2.2 ¿specific inspection¿): any type of deformation, dents, cracks, deep scratches, defects in and at the insulation, dirt, corrosion, and labelling visibility.Olympus will continue to monitor field performance for this device.
|