• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIRPTR 2.3 PK SUTURE ANCHOR W/ ULTRAB; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIRPTR 2.3 PK SUTURE ANCHOR W/ ULTRAB; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72201541
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Perforation (2001); Device Embedded In Tissue or Plaque (3165)
Date of Event 09/02/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: case (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, the thread of the bioraptor anchor broke when the surgeon tied the thread and pulled the suture for tying the knot.Part of the broken anchor remained inside the patient bone and the back-up was used in the originally drilled bone hole.The procedure was completed with non-significant delay using a smith and nephew back up device.No further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that a material certification of analysis is required with each lot.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.H11: corrected information in h6 (health effect - clinical & impact codes).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIRPTR 2.3 PK SUTURE ANCHOR W/ ULTRAB
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17805199
Report Number1219602-2023-01839
Device Sequence Number20515037
Product Code MBI
UDI-Device Identifier03596010595072
UDI-Public03596010595072
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K152566
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201541
Device Lot Number2099337
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2023
Supplement Date Received by Manufacturer01/10/2024
Initial Report FDA Received Date09/24/2023
Supplement Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age18 YR
Patient SexMale
-
-