| Catalog Number |
72201541 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Perforation (2001); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
09/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, the thread of the bioraptor anchor broke when the surgeon tied the thread and pulled the suture for tying the knot.Part of the broken anchor remained inside the patient bone and the back-up was used in the originally drilled bone hole.The procedure was completed with non-significant delay using a smith and nephew back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that a material certification of analysis is required with each lot.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.H11: corrected information in h6 (health effect - clinical & impact codes).
|
| |
|
Search Alerts/Recalls
|