| Catalog Number |
64811111 |
| Medical Device Problem Codes |
Fracture (1260); Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
08/31/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hinged knee was revised after patient complaint of thigh pain.Surgeon originally suspected femoral loosening.However, intra-op it was observed that the lateral condyle of the femoral component broke off of the implant and was loose in the joint.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding crack/fracture and loosening involving a mrh femoral component was reported.The event was confirmed for crack/fracture.Method & results: product evaluation and results: the reported device was not returned; however, photographs were provided for review.The photographs show a recently explanted femoral component with fractured condyle.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient had pain; intraop it was observed that the implant was fractured and loose.The reported device was not returned; however, photographs were provided for review.The photographs show a recently explanted femoral component with fractured condyle.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hinged knee was revised after patient complaint of thigh pain.Surgeon originally suspected femoral loosening.However, intra-op it was observed that the lateral condyle of the femoral component broke off of the implant and was loose in the joint.
|
| |
|