It was reported by the distributorship that the products were found to be nonconforming, found debris in the sterile package.The event occurred outside of surgery.There was no reported patient harm, or delay.Due diligence is complete, no further information is available at this time.
|
|
This event has been recorded by zimmer biomet under cmp-(b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d2, d4, g1, g3, h1, g6, h2, h3, h4, h6, h8 and h10 visual inspection confirmed there was debris inside the sealed package of 1 of the tips.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|