| Catalog Number |
72203981 |
| Medical Device Problem Codes |
Break (1069); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a rotator cuff repair, the threads of the healicoil anchors peeled on insertion into the humeral head, there was separation of the threads from the core anchor.All the anchor material was removed from the patient with the help of forceps.Non-significant surgical delay was reported, and the procedure was finished with a smith and nephew back up device used in the originally drilled bone hole, no void left in the patient.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the raw material strength requirements and storage requirements specified and each lot must be accompanied by a material certificate of analysis.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|