| Model Number |
02-01804 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Melted (1385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The device returned with physical damage on the top of the monitor.The damage appeared to be caused by overheat.The patient never reported any technical or physical failures of the monitor during use.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned with physical damage near the camera.The patient did not report any technical issues or malfunctions during use.The device was investigated.Engineering evaluation was able to confirm device melt.No internal device was observed during evaluation.It is most probable that the device melt was caused by an external heat source.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d4.
|
| |
|