| Lot Number |
0029596705 |
| Medical Device Problem Codes |
Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 30-aug-2022.It was reported that device damage occurred.A 4.00 x 48mm synergy xd drug-eluting stent was selected for treatment, but the device was noted to be damaged.No patient complications were reported.However, returned device analysis revealed stent detachment.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: synergy xd mr us 4.00 x 48mm stent delivery system (sds) was returned for analysis.Visual/tactile, microscopic and dimensional analyses were performed on the device.The stent was returned detached from the device.Damage along the entire length of the stent was noted with the stent struts pulled and stretched evidencing excessive tensile force applied to the stent.A stent od reading couldn't be obtained.A review of the manufacturing stent profile data was performed and the stent od at the time of manufacture was 0.0484 inches which is within specification.Balloon cones were reviewed where it was subjected to positive pressure.No issues were identified with the hypotube shaft.No issues were identified with the outer/mid-shaft sections or the inner lumen of the device.Bumper tip showed no signs of distal tip damage.
|
| |
|