• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TUBES BLUSELECT SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TUBES BLUSELECT SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/870/075
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other text: b3: month and year of event have been provided; day is unknown.D4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the pilot balloon was replaced in the morning, and in the afternoon of the same day.It was discovered that the pilot balloon had fallen off, and use was discontinued.Adverse patient effects are unknown.
 
Additional Manufacturer Narrative
Device evaluation: one device was returned for investigation.Under visual inspection the inflation line was found detached from the pilot balloon as reported by customer.The complaint was confirmed.This was listed as a supplied item fault.This issue will continue to be monitored and further actions taken accordingly.No lot number was provided therefore no device history report (dhr) review could be completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT SUCTIONAID
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17813725
Report Number3012307300-2023-09225
Device Sequence Number6727368
Product Code BTO
UDI-Device Identifier15019517076219
UDI-Public15019517076219
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number101/870/075
Device Catalogue Number101/870/075CZ
Was Device Available for Evaluation? No
Date Returned to Manufacturer09/07/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2023
Supplement Date Received by Manufacturer10/09/2023
Initial Report FDA Received Date09/25/2023
Supplement Report FDA Received Date11/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-