• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL Maestro Cochlear Implant System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL Maestro Cochlear Implant System Back to Search Results
Model Number Mi1210 SYNCHRONY ST
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
Affected channels were reported.Re-implantation is considered.
 
Additional Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode as might be caused by an external mechanical impact appears likely.However, to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
 
Event or Problem Description
Affected channels were reported.Re-implantation is considered.No date has been scheduled yet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL Maestro Cochlear Implant System
Common Device Name
Cochlear Implant
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
+43-(577)88
MDR Report Key17815742
Report Number9710014-2023-00850
Device Sequence Number14992862
Product Code MCM
UDI-Device Identifier09008737366193
UDI-Public(01)09008737366193
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/08/2025
Device Model NumberMi1210 SYNCHRONY ST
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/23/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2023
Supplement Date Received by Manufacturer09/24/2023
09/24/2023
Initial Report FDA Received Date09/26/2023
Supplement Report FDA Received Date10/12/2023
07/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age4 YR
Patient SexMale
-
-