| Model Number |
CRPLUS |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer will be provided with a replacement device.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that their device would not power on, when attempted.As a result, defibrillation therapy would not be available, if needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker product analysis center (pac) evaluated the customer's device and verified the reported issue.It was observed that the on/off button was disconnected from its hinges.After reconnecting the button, proper device operation was observed.The device was archived by stryker.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that their device would not power on, when attempted.As a result, defibrillation therapy would not be available, if needed.There was no patient use associated with the reported event.
|
| |
|