| Model Number |
ENSITE-DWS-1.1 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, there was not output from the remote video and the procedure was cancelled.The dws was rebooted which restored remote video, but then there were amplifier communication issues.After restarting both the dws and amplifier several times, communication was not established, and the case was cancelled.The patient was rescheduled for oct.6.The device is still being used at the facility.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported output issue remains unknown.
|
| |
|