| Model Number |
22550A |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
Superficial (First Degree) Burn (2685)
|
| Date of Event |
08/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported during a direct current cardioversion to treat an atrial flutter there was a spark on the anterior left chest pad.The patient converted to sinus rhythm after 1 shock at 250 joules and the pad was removed from the patient's chest.Two thermal burns approximately 1/2"x2" were found.The patient received superficial (first degree) burns.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product, or the subassemblies that could have led to the reported condition.No samples have been received.Without samples it is not possible to determine root causes or to determine if they could be manufacturing related.No corrective or preventative actions will be taken at this time.This complaint will be used for tracking and trending purposes.
|
| |
|