• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 03666794001
Medical Device Problem Codes Display Difficult to Read (1181); Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.
 
Event or Problem Description
We received an allegation of a display issue with coaguchek xs meter serial number (b)(6).The customer stated the meter display is faint and difficult to read.There was no misinterpretation of results.
 
Additional Manufacturer Narrative
The meter was returned for investigation.The returned meter did not power on.The meter was optically inspected and contamination from fluid ingress was observed on the printed circuit board.The contamination caused a power interruption which prevents the meter from powering on.Additionally, the contamination can cause faded/missing segments.The cause of the issue is contamination due to user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17817240
Report Number1823260-2023-03102
Device Sequence Number14992781
Product Code GJS
UDI-Device Identifier04015630945689
UDI-Public04015630945689
Combination Product (Y/N)Y
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/05/2023
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date09/26/2023
Supplement Report FDA Received Date10/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
-
-