| Catalog Number |
03666794001 |
| Medical Device Problem Codes |
Display Difficult to Read (1181); Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation of a display issue with coaguchek xs meter serial number (b)(6).The customer stated the meter display is faint and difficult to read.There was no misinterpretation of results.
|
| |
|
Additional Manufacturer Narrative
|
|
The meter was returned for investigation.The returned meter did not power on.The meter was optically inspected and contamination from fluid ingress was observed on the printed circuit board.The contamination caused a power interruption which prevents the meter from powering on.Additionally, the contamination can cause faded/missing segments.The cause of the issue is contamination due to user error.
|
| |
|