|
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the emitter had a red x through it.Emitter details report: axiem controller not connected message appeared.The site noticed that the emitter was not working. there was no patient involvement.Troubleshooting was performed, the site could not hear the emitter.There was a number on the axiem controller that looked to be a number 8.The system was restarted, the axiem control was unplugged for 30-60 seconds, and the axiem was plugged back in.An orange dotted line was on the setup equipment.The fans ran when the axiem was plugged back in, but the axiem controlled not connected message still appeared.
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9733320, serial/lot #: unknown, h3) no parts have been received by the manufacturer for evaluation.Codes: b17, c20, d15.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|