• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOPHYSA PRESSIO ICP MONITORING KIT, PARENCHYMAL TUNNELLING; ICP MONITORING CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOPHYSA PRESSIO ICP MONITORING KIT, PARENCHYMAL TUNNELLING; ICP MONITORING CATHETER Back to Search Results
Model Number PSO-PT
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
After approx 24 hours an error code e001 came up on the monitor.The hospital staff tried changing the console and cables all to no effect.The catheter was removed and a new one inserted.
 
Additional Manufacturer Narrative
Error e001 confirmed on monitor.Abnormal extension of the pa tube over approximately 2 cm which caused the cable to break in the pa tube.User error confirmed.
 
Event or Problem Description
After approx 24 hours an error code e001 came up on the monitor.The hospital staff tried changing the console and cables all to no effect.The catheter was removed and a new one inserted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSIO ICP MONITORING KIT, PARENCHYMAL TUNNELLING
Common Device Name
ICP MONITORING CATHETER
Manufacturer (Section D)
SOPHYSA
5 rue guy moquet
orsay, 91400
FR  91400
Manufacturer (Section G)
SOPHYSA
rue sophie germain
technopôle temis
besancon, 25000
FR   25000
Manufacturer Contact
marion prudhomme
5 rue guy moquet
orsay, 91400
FR   91400
MDR Report Key17817607
Report Number3001587388-2023-23419
Device Sequence Number14986990
Product Code GWM
UDI-Device Identifier03760124130782
UDI-Public(01)03760124130782(11)221014(17)270901
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K162108
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial,Followup
Report Date (Section B) 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPSO-PT
Device Catalogue NumberPSO-PT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/21/2023
Supplement Date Received by Manufacturer09/21/2023
Initial Report FDA Received Date09/26/2023
Supplement Report FDA Received Date01/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-