|
|
| Model Number |
KD-645L |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that during peroral endoscopic myotomy procedure, the distal part of the single use electrosurgical knife was burnt and fell into the patient's esophagus.The fallen part was removed by forceps; however, it was so small (shrunk); it was not possible to preserve it.As reported, the defect was clearly visible on the knife.The procedure was a bit longer due to the retrieval.The procedure was completed using a similar device.The device was inspected before use and no abnormalities were observed.There were no reports of patient or user harm associated with this event.
|
| |
|
Event or Problem Description
|
|
The tip fall off at the beginning of the myotomy, right after the incision.But it can be said that it started to appear already during the incision.The power settings in (poem) peroral endoscopic myotomy have been the same in the past couple of years.The user has used vio300d in the past.Co2 gas was delivered for the procedure.During the myotomy spraycoag mode was used.The user used tunneling for spraycoag mode.No video can be provided for the procedure in question.The tip dropout occurred during spraycoag.The order of the poem procedure flow when the tip dropout occurred was: standard poem , injection, incision, tunneling.No mechanical load on the tip during that incision.The tip dropout was gradual.The triangular tip started to deform during the incision.User used the second mode for the incision.Endo cut mode: effect 2 with upmax:770, and age cur 1 with interval cut 6.Spray coagulation mode: effect 2 with upmax 4300vp, and max watt 50.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation with correction to the initial with information inadvertently left out.Additionally, to provide information provide through follow up ((b)(4)).The device was returned to olympus for inspection, and the customer's complaint/reportable malfunction was confirmed.The complete evaluation results are as followed: location of the cutting knife breakage was at 3.7mm away from the distal cover.It was discovered that the cutting knife was broken near the triangle tip.The cutting knife was scorched.The foreign substances had adhered around the distal end.The distal end of the item was inspected under a microscope and detected that the surface of the damaged cutting knife.Plus, the surface was melted.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the suggested event "distal part of knife was burned and fell into patient" occurred due to the following: 1.) during the procedure, an electrical discharge might have occurred due to one of the following factors.¿the setting of the electrosurgical unit was coagulation mode when the device was used for the procedure.¿the activation time was too long.2.) an electrical discharge occurred, and the part of the cutting wire became hot instantly.As a result, the strength of the cutting knife decreased.Also, the cutting wire was scorched.3.) during tissue incision, a load was applied to the cutting knife which has the reduced strength.This might have caused the cutting knife to break and fell off.However, the root cause of the suggested event could not be determined.The event can be prevented by following the instructions for use which state: "use this instrument only in combination with products recommended by olympus.If combined with products not recommended by olympus, patient injury caused by increase in patient leakage current, operator injury, malfunction or equipment damage may result." "do not use this instrument in an output higher than the rated high-frequency voltage in the table on page 4.This could cause patient, operator, or assistant injury, such as thermal injury.It could also damage the endoscope, instrument." "8.2 specifications" "during treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.Should deformation or break of the cutting knife and the triangle tip be detected during use, immediately shut off the power supply, discontinue the procedure, pull the slider and, withdraw the endoscope from the patient with the cutting knife retreated in the coated outer tube.Do not continue using an abnormal electrosurgical knife to prevent perforation or bleeding.If the cutting knife and/or the triangle tip is detached, be sure to collect it using a grasping forceps." "always operate the electrosurgical unit at the minimum output level and for the minimum time necessary to successfully complete the procedures.An excessive output level and time may result in patient injury, such as perforation, bleeding, or mucous membrane damage.Application of high-voltage waveforms for extended periods increases the likelihood that the cutting knife and the triangle tip may break.When a high-voltage waveform has to be used, minimize the duration of current application.-electrosurgical unit: voltage intensities of various waveforms soft coagulation < cut / pulse cut < forced coagulation < spray coagulation" "do not activate output continuously but activate it intermittently.Continuous activation may result in patient injury, such as bleeding, tissue damage, or thermal injury of non-target tissue.Breakage or deformation of the triangle tip, and cutting knife may likely occur." "stop activating output immediately if the triangle tip and the cutting knife are found to turn red while activating output.Keeping output activated while the triangle tip and the cutting knife are red may result in thermal injury.Detachment of the triangle tip and deformation/break of the triangle tip, and cutting knife may also occur." "be careful not to use excessive force when removing tissue attached to the cutting knife and the triangle tip.When the distal end is subjected to excessive force, for example, when forcibly scraping the cutting knife with tweezers, etc.Or when extending and retracting the cutting knife abruptly, and continuously, it may break the cutting knife and the triangle tip." olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide further testing results from the device evaluation and legal manufacturer¿s investigation findings.Further analysis of the fractured area of the subject device and replication testing was conducted by olympus.The results are as follows: the location of the cutting knife breakage was 4.5 mm away from the distal cover.It was noted that the cutting knife was broken near the triangle tip.Since the triangle tip was not returned to olympus for evaluation, the device was not inspected, and the details including the shape are unknown.Based on the further investigation results, it was determined that the triangle tip, which had become brittle due to melting, experienced brittle fracture under mechanical loading.Also, the shape of the distal end of the triangle tip likely deformed due to mechanical loading.The likely mechanism causing the reported event is as follows: 1) during the procedure, the amount of discharge increased due to the following factors: a.The cutting knife was energized while away from the tissue and then approached the tissue.As a result, the voltage increased, leading to an increase in the amount of electrical discharge.B.The output activation time was too long.2) an electrical discharge occurred, and the part of the cutting knife became hot instantly.As a result, the strength of the cutting knife decreased.Also, the cutting wire was scorched.3) during handling such as retracting the cutting knife, a load was applied to the cutting knife which has reduced its strength.This might have caused the cutting knife to break and detach.However, the root cause of the reported event could not be determined.This supplemental report includes a additional information to h6 (investigation findings) olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|