| Catalog Number |
4825202 |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process, a follow-up report will be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in b.5, h.6 and h.10.Investigation: the run data file (rdf) was analyzed for this event.Run data file analysis did not show a conclusive root cause for the elevated wbc content in the platelet product reported for this collection.Alerts that are known to contribute to wbc contamination were not generated in this run data file.As the trima accel system cannot count the cells entering the platelet bag, the rbc detector signals must see a significant change in the reflectance values to notify the operator of an lrs chamber saturation and to count the wbcs in the platelet product.In this case, the signals did not indicate that wbcs were escaping the lrs chamber.Therefore, the run data file reported that the platelet product could be labeled as leukoreduced.Although the trima accel system has several methods for detection of possible wbc contamination, it is possible that some events may elude the detection capability of the trima accel system.It is also possible that this leukoreduction failure is donor related.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Investigation is in process, a follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6) the platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide investigation: the run data file (rdf) was analyzed for this event.Run data file analysis did not show a conclusive root cause for the elevated wbc content in the platelet product reported for this collection.Alerts that are known to contribute to wbc contamination were not generated in this run data file.As the trima accel system cannot count the cells entering the platelet bag, the rbc detector signals must see a significant change in the reflectance values to notify the operator of an lrs chamber saturation and to count the wbcs in the platelet product.In this case, the signals did not indicate that wbcs were escaping the lrs chamber.Therefore, the run data file reported that the platelet product could be labeled as leukoreduced.Although the trima accel system has several methods for detection of possible wbc contamination, it is possible that some events may elude the detection capability of the trima accel system.It is also possible that this leukoreduction failure is donor related.Investigation is in process, a follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in h.6 and h.10.Investigation: the run data file (rdf) was analyzed for this event.Run data file analysis did not show a conclusive root cause for the elevated wbc content in the platelet product reported for this collection.Alerts that are known to contribute to wbc contamination were not generated in this run data file.As the trima accel system cannot count the cells entering the platelet bag, the rbc detector signals must see a significant change in the reflectance values to notify the operator of an lrs chamber saturation and to count the wbcs in the platelet product.In this case, the signals did not indicate that wbcs were escaping the lrs chamber.Therefore, the run data file reported that the platelet product could be labeled as leukoreduced.Although the trima accel system has several methods for detection of possible wbc contamination, it is possible that some events may elude the detection capability of the trima accel system.It is also possible that this leukoreduction failure is donor related.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Further evaluation of this event has determined that the device did not cause or contribute to a death or serious injury, nor is there a likely potential for death or serious injury associated with this event based on additional investigational information.During customer follow-up, it was confirmed that the wbc count was below ¿¿¿5x10^6 per transfusable unit.Root cause: run data file analysis did not show a conclusive root cause for the elevated wbc content in the platelet product reported for this collection.Alerts that are known to contribute to wbc contamination were not generated in this run data file.As the trima accel system cannot count the cells entering the platelet bag, the rbc detector signals must see a significant change in the reflectance values to notify the operator of an lrs chamber saturation and to count the wbcs in the platelet product.In this case, the signals did not indicate that wbcs were escaping the lrs chamber.Therefore, the run data file reported that the platelet product could be labeled as leukoreduced.Although the trima accel system has several methods for detection of possible wbc contamination, it is possible that some events may elude the detection capability of the trima accel system.It is also possible that this leukoreduction failure is donor related.No further reporting will be provided as this does not represent a reportable event.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: (b)(6) the platelet collection set is not available for return because it was discarded by the customer.
|
| |
|