Catalog Number TFGT-25A
Medical Device Problem Codes Incomplete Coaptation (2507); Patient-Device Incompatibility (2682); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/29/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2019, a 23mm trifecta gt valve was implanted in a patient.On (b)(6) 2023, the ultrasonic echocardiography showed pannus on the valve.On (b)(6) 2023, the valve got explanted.There was impaired movement of the valve cusp due to pannus, and there was incomplete coaptation was due to separation of cusp.A new 23mm non-abbott valve was chosen and implanted.The patient was reported stable.
 
Additional Manufacturer Narrative
Explant due to an ultrasonic echocardiography showing pannus was reported.The investigation found that leaflets 1 and 2 were torn.There was fibrous pannus ingrowth on the inflow surface of all leaflets and on the outflow of leaflet 1.There was no inflammation or significant calcifications.The device history record was also reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.In the absence of any calcification or evidence for infection, the reported event is consistent with a non-calcific leaflet tear.Non-calcific leaflet tear is a form of structural valve deterioration (svd), which is a well-known complication from valve replacement surgery.A non-calcific leaflet tear is commonly attributed to increased operational leaflet stress but may also be related to biological factors which result in tissue degeneration characterized by loss of collagen.In addition, the pannus noted could have increased stress on adjacent leaflets and created an unbalanced stress relief distribution between all leaflets during coaptation, which may lead to leaflet tears and reduced durability.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIFECTA¿ GT VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17818803
Report Number2135147-2023-04182
Device Sequence Number6205433
Product Code LWR
UDI-Device Identifier05415067018236
UDI-Public05415067018236
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P100029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2022
Device Catalogue NumberTFGT-25A
Device Lot Number6544254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/01/2023
Supplement Date Received by Manufacturer12/11/2023
Initial Report FDA Received Date09/26/2023
Supplement Report FDA Received Date12/11/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/25/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age61 YR
Patient SexMale
Patient Weight57 KG
-
-