SYNTHES GMBH TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 04.043.145S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2: additional procode: hwc d9: complainant part is not expected to be returned for manufacturer review/investigation.D10: date of concomitant therapy is (b)(6) 2023.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device product code: 04.043.145s-us lot number: 589p209 it was electronically reviewed and no non conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 28/01/2022.Manufacturing site: jabil bettlach.Expiry date:31/12/2031 device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2023, during a tibial nail case, the surgeon changed to a smaller suprapatellar sleeve as they were having difficulty getting to the starting point.The surgeon was eventually able to get to the starting point and proceeded with the surgery.When putting in the proximally locking screws, the drill bit in the s1 screw hole hit the nail.The surgeon was unable to get the bit past the nail and so chose other screw holes where the screws were successfully inserted.The screw hole alignments were checked prior to nail insertion and they all lined up, the surgeon felt that the aiming arm was not properly latched.It was discovered at the end of the surgery that the spring on the latch of the aiming arm was not working.The procedure was completed successfully with no adverse patient outcome.There was a surgical delay of 45 minutes.No fragments were generated.No additional information is available to report.This report is for a tibial nail-advanced / 9mm 375mm / sterile.This is report 2 of 3 for pc (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17819662
Report Number8030965-2023-12147
Device Sequence Number11348527
Product Code JDS
UDI-Device Identifier10886982297655
UDI-Public(01)10886982297655
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K201336
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.043.145S
Device Lot Number589P209
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/08/2023
Initial Report FDA Received Date09/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AIMING ARM/ RADIOLUCENT; PROTEC SLV/ FLEX/ LG FOR NL 8-11 / -S; UNK - DRILL BITS: TRAUMA; UNK - SCREWS: LOCKING
Patient Age26 YR
Patient SexFemale
-
-