| Model Number |
100-9810 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent an explant procedure to explant the indirect decompression (id) spacer.The patient had gotten into a golf cart accident and broken one of the legs off the spacer.Although it was not easy to get the spacer out, the physician successfully explanted the spacer and the broken leg component.The patient was reimplanted with a new spacer at the same level and an additional new spacer was added at l 3-4 lumbar vertebrae.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an explant procedure to explant the indirect decompression (id) spacer.The patient had gotten into a golf cart accident and broken one of the legs off the spacer.Although it was not easy to get the spacer out, the physician successfully explanted the spacer and the broken leg component.The patient was reimplanted with a new spacer at the same level and an additional new spacer was added at l 3-4 lumbar vertebrae.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned spacer was analyzed and visual examination confirmed the event of the patient having a damaged spacer due to a golf cart accident.The returned implant revealed a fractured cam lobe, and tool damage to the spindle, spindle cap, and body of the implant.The damage to the implant was sufficient to prevent functional testing.A labeling review was performed and it did not reveal any anomalies as the instructions for use (ifu) states that device dislodgement, particularly if the patient experiences a traumatic event postoperatively and loosening, fatigue, deformation, breakage or disassembly of the implant, which may require another operation to remove the implant and may require another method of treatment; as well as deformation, breakage or disassembly of the implant, and spinous process fracture are among other risks associated with the procedure.The reported event of the patient having a damaged spacer due to a golf cart accident was confirmed.Visual inspection revealed a fractured cam lobe, and tool damage to the spindle, spindle cap and body of the implant.The damage to the implant was sufficient to prevent functional testing.However, a labelling review found that the reported event is a known risk associated with the use of lumbar spine implants and associated instruments.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an explant procedure to explant the indirect decompression (id) spacer.The patient had gotten into a golf cart accident and broken one of the legs off the spacer.Although it was not easy to get the spacer out, the physician successfully explanted the spacer and the broken leg component.The patient was reimplanted with a new spacer at the same level and an additional new spacer was added at l 3-4 lumbar vertebrae.
|
| |
|