|
The device was not returned to olympus for evaluation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The subject device was manufactured on september, 2022 based on the provided 3 digit lot information but the exact manufacture date could not be identified since the subject device was not returned.Based on the results of the investigation, a definitive root cause cannot be identified.It is likely the malfunction is a result of a force to perform tissue incision being applied to the tip, high heat caused by spark discharge being locally applied to the tip and the electrode causing the adhesive strength of the adhesive agent affixing the cutting knife and the tip to decrease, or a spark discharge between the tissue and the electrode occurred during output activation caused by the high-frequency output being set too high, the output setting for activation being too high, the electrosurgical unit was used in the coagulation mode, or the output activation time was too long.This issue is addressed in the instructions for use (ifu) under section 10.3 operation ¿ cutting.-when the electrosurgical unit is used in the coagulation mode, crack/detachment of the tip or deformation/break of the cutting knife or the electrode could occur, for example when the high-frequency output is set too high or the length of the contact between the cutting knife and tissue is too short.During treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.Should cracks or detachment of the tip or deformation/break of the cutting knife be detected during use, immediately shut off the power supply, discontinue the procedure, pull the slider and withdraw the endoscope from the patient with the cutting knife retreated in the insertion portion of this instrument.Do not continue using an abnormal electrosurgical knife to prevent perforation or bleeding.If the tip or cutting knife is detached be sure to collect it using grasping forceps.-aspirate fluids such as mucus that adhere to the electrode and/or the cutting knife, outer sheath and body cavity tissues.Patient injury such as punctures, hemorrhages, mucous membrane damage and thermal injury of tissue could result if output is activated when in contact with these adhering fluids.When current is discharged while the cutting knife is being separated from the mucosa under wet situation, it may break the cutting knife or crack the tip.-always operate the electrosurgical unit at the minimum output level and for the minimum time necessary to successfully complete the procedures.Excessive output level and time may occur in patient injury, such as perforations, bleeding or mucous membrane damage.Application of high-voltage waveforms for extended periods increase the likelihood that it may break the cutting knife or crack the tip.When a high-voltage waveform has to be used, minimize the duration of current application.-electrosurgical unit: voltage intensities of various waveforms soft coagulation < cut / pulse cut < forced coagulation -do not activate output continuously but activate it intermittently.Continuous activation may cause patient injury, such as bleeding, mucous membrane damage, or thermal injury of non target tissue.Crack/detachment of the tip or deformation/break of the cutting knife or electrode may also occur.-do not activate output when the electrode and cutting knife are not in contact with the target tissue.Crack/detachment of the tip or deformation/break of the cutting knife or electrode may occur.Olympus will continue to monitor the field performance of this device.
|