• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TENEX HEALTH, INC. TX1 TISSUE REMOVAL SYSTEM - MICROTIP; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TENEX HEALTH, INC. TX1 TISSUE REMOVAL SYSTEM - MICROTIP; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 554-1003-001
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
During a procedure with the tenex system, the microtip needle separated from the rest of the hand piece.There were no patient complications.
 
Additional Manufacturer Narrative
Corrections were made to the device brand name, model number and labeling information.The device was returned and evaluated.It was confirmed that the microtip needle had separated from the rest of the handpiece.Due to the state in which the device was received, functional testing could not be performed.The fracture site did not immediately indicate a specific cause for the failure.Disassembly of the device did not reveal any further anomalies or defects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TX1 TISSUE REMOVAL SYSTEM - MICROTIP
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630 8123
Manufacturer (Section G)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630
Manufacturer Contact
david vancelette
26902 vista terrace
lake forest, CA 92630
9492388220
MDR Report Key17821852
Report Number1000135560-2023-00044
Device Sequence Number6124787
Product Code LFL
UDI-Device Identifier00857156006215
UDI-Public(01)00857156006215(10)0232301(17)250122
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K123640
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/23/2025
Device Model Number554-1003-001
Device Catalogue Number554-1003-001
Device Lot Number1142301
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2023
Supplement Date Received by Manufacturer08/31/2023
Initial Report FDA Received Date09/26/2023
Supplement Report FDA Received Date10/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight78 KG
-
-