• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL OY VITAL SIGNS¿ LEG ATTACHMENT PAD; ELECTRODE, ELECTROCARDIOGRAPH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL OY VITAL SIGNS¿ LEG ATTACHMENT PAD; ELECTRODE, ELECTROCARDIOGRAPH Back to Search Results
Model Number LEG ATTACHMENT PAD FOR COROMETRICS FETAL SCALP ELECTRODE CABLE
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event or Problem Description
The customer reported to vyaire medical that the leg attachment pad for corometrics fetal scalp electrode cable does not stay securely and does not make good contact during patient use.The device is easily detached, which results in inaccurate fetal heart rate reading.Additionally, there is no information regarding patient harm.
 
Additional Manufacturer Narrative
Result of investigation: the device history record (dhr) for lot 314665 has been reviewed, paying close attention to the foam and snap lots as well as in process inspection records such as ec12 testing.A part from ncr-ca-hl-22-007 affecting the stud lot, no abnormalities were found and products were manufactured to specification.The adhesion of the product has been reviewed as a part of capa investigation capa-000001058.All lots have been within specification.Because of this complaint and others relating to the lack of adhesiveness of this product, we have determined that the other potential root cause may be related to the adhesion of this product not being identical to the previous product.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITAL SIGNS¿ LEG ATTACHMENT PAD
Common Device Name
ELECTRODE, ELECTROCARDIOGRAPH
Manufacturer (Section D)
VYAIRE MEDICAL OY
kuortaneenkatu 2
etela-suomen laani
helsinki 00510
FI  00510
Manufacturer (Section G)
GRAPHIC CONTROLS ACQUISITION CORP.
400 exchange street
buffalo NY 14204
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
2402760001
MDR Report Key17822536
Report Number3010838917-2023-00085
Device Sequence Number14063607
Product Code DRX
UDI-Device Identifier10190752188181
UDI-Public(01)10190752188181(10)314665
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLEG ATTACHMENT PAD FOR COROMETRICS FETAL SCALP ELECTRODE CABLE
Device Catalogue Number2464AAO-20
Device Lot Number314665
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2023
Supplement Date Received by Manufacturer10/15/2023
Initial Report FDA Received Date09/26/2023
Supplement Report FDA Received Date10/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-