The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the lg connector fluid damage, the remote control buttons cracked, the lcb distal cover glass cracked, the lg connector corroded, the bending rubber leak, the remote control buttons leak, the root brace rubber (insertion flexible tube) cut, the suction channel kink, the insertion flexible tube dented, and the bending rubber pin hole; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|