| Catalog Number |
A16MA02101512 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Pain (1994); Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that hedstroem m-access 21mm 015 reportedly appears that this file had broken in the root canal during treatment.Patient still felt pain four months after the root canal treatment.Examination revealed a 3 mm broken file in the root canal.Root canal was repeated along with anti-inflammatory treatment and after 15 days the patient's symptoms improved.Treatment ended.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1716602).Root causes are not identified.We will track this kind of event and monitor the trend.
|
| |
|