BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2023.It was reported that the tip of the guidewire was stretched during use.A comet ii pressure guidewire was used for cardiac intervention.During the procedure, the tip of the guidewire was stretched, and the tip could not be shaped.After repeated attempts, it still could not reach the distal end of the blood vessel.Another pressure guidewire was used, and the procedure was completed.No patient complications were reported, and patient was stable post procedure.However, device analysis revealed that the tip was stretched for 7mm in length and detached at an unknown location.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the complaint device was received for product analysis.Returned product consisted of an ffr comet pressure wire with the occ cable.The occ cable, tip, and device shaft were visually and microscopically examined for damage or any irregularities.Inspection of the device revealed that the tip was stretched for 7mm in length and detached at an unknown location (due to the stretching of the tip, the detachment location cannot be measured).The ends of the detachment appeared to be smooth with material pushed up at an edge.The detached portion of the tip was not returned for analysis.The occ cable was then connected to the bench top testing equipment and the coefficient values were confirmed to be programmed.No other issues were identified during the product analysis.Product analysis confirmed the reported event, as the tip was stretched and detached.
 
Event or Problem Description
Reportable based on device analysis completed on 01sep2023.It was reported that the tip of the guidewire was stretched during use.A comet ii pressure guidewire was used for cardiac intervention.During the procedure, the tip of the guidewire was stretched, and the tip could not be shaped.After repeated attempts, it still could not reach the distal end of the blood vessel.Another pressure guidewire was used, and the procedure was completed.No patient complications were reported, and patient was stable post procedure.However, device analysis revealed that the tip was stretched for 7mm in length and detached at an unknown location.It was further reported that the tip of the device was stretched and there is no detached portion.Also, no fracture was noted.Device was completely removed from the patient body.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the complaint device was received for product analysis.Returned product consisted of an ffr comet pressure wire with the occ cable.The occ cable, tip, and device shaft were visually and microscopically examined for damage or any irregularities.Inspection of the device revealed that the tip was stretched for 7mm in length and detached at an unknown location (due to the stretching of the tip, the detachment location cannot be measured).The ends of the detachment appeared to be smooth with material pushed up at an edge.The detached portion of the tip was not returned for analysis.The occ cable was then connected to the bench top testing equipment and the coefficient values were confirmed to be programmed.No other issues were identified during the product analysis.Product analysis confirmed the reported event, as the tip was stretched and detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Common Device Name
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17824753
Report Number2124215-2023-52743
Device Sequence Number2440436
Product Code DXO
UDI-Device Identifier08714729904397
UDI-Public08714729904397
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151610
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2023
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0028488566
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/01/2023
Supplement Date Received by Manufacturer10/07/2023
Initial Report FDA Received Date09/27/2023
Supplement Report FDA Received Date10/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age64 YR
Patient SexMale
Patient Weight79 KG
-
-