| Catalog Number |
1651-38 |
| Medical Device Problem Codes |
Deflation Problem (1149); Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for investigation.The balloon would not inflate during testing.The catheter was cut and inspected; no blockage or foreign material was observed blocking the inflation lumen and fluid was still unable to pass through the catheter.The y-connector was observed and fluid was unable to pass through the y-connector indicating a blockage was present in the y-connector.The device was cut approximately half way down the length of the y-connector and fluid was then able to pass through the inflation lumen.While the reported issue was confirmed it could not be determined if the blockage occurred during manufacturing or while handling the device.No inflation issues were noted during manufacturing that would indicate a blockage was present during quality inspection.The lot history record for the device was reviewed, no issues were found during manufacturing or packaging that would cause or contribute to the reported event.We have not received any similar complaints related to this lot number.
|
| |
|
Event or Problem Description
|
|
The balloon did not work unless it was inflated and deflated much stronger than usual during inflation and deflation check.Therefore, the catheter was replaced with a new one.The operation was successfully completed afterwards.No injury was reported to the patient.
|
| |
|