MAUDE Adverse Event Report: ARTHREX, INC. LOW PRO SCRW,TI,6.7X 55MMCANN,18MM THD; BONE FIXATION PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. LOW PRO SCRW,TI,6.7X 55MMCANN,18MM THD; BONE FIXATION PLATE Back to Search Results
Model Number LOW PRO SCRW,TI,6.7X 55MMCANN,18MM THD
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a surgery the thread of the screw has loosened during insertion.There was no harm for patient, operator or third party reported.No further information received.Update avoe 31-aug-2023: it was confirmed that the error occurred during an osg arthrodesis surgery and all fragments were retrieved.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during insertion of the screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOW PRO SCRW,TI,6.7X 55MMCANN,18MM THD
Common Device Name
BONE FIXATION PLATE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17825154
Report Number1220246-2023-07996
Device Sequence Number6161789
Product Code HRS
UDI-Device Identifier00888867056022
UDI-Public00888867056022
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K141478
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLOW PRO SCRW,TI,6.7X 55MMCANN,18MM THD
Device Catalogue NumberAR-8967-1855
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2023
Supplement Date Received by Manufacturer02/17/2025
Initial Report FDA Received Date09/27/2023
Supplement Report FDA Received Date02/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-